05-December-2025
PL40
50
To define the standard procedure for checking and verifying the output of a ratiometric pressure transmitter by applying known pressure values and measuring the corresponding output voltage.
This procedure applies to pressure transmitters with ratiometric voltage output, where the output voltage is proportional to the supply voltage.
The output voltage of the transmitter is a fixed percentage of the supply voltage.
If the supply voltage changes, the output voltage also changes proportionally.
Therefore:
Knowing and measuring the supply voltage is mandatory before checking output accuracy.
Regulated DC power supply (as per transmitter specification)
Digital multimeter (for voltage measurement)
Pressure source / pressure calibrator (–1 to +5 bar)
Test leads and suitable fittings
Transmitter must be properly mounted or securely connected.
Correct supply voltage must be applied as per datasheet.
Pressure source must be accurate and leak-free.
Ambient temperature conditions should be stable.
Connect the supply voltage to the pressure transmitter.
Connect the multimeter to the output terminals in voltage measurement mode.
Connect the pressure source to the pressure port of the transmitter.
Switch ON the power supply.
Measure and record the actual supply voltage using a multimeter.
Ensure the supply voltage is within the specified range.
Keep the transmitter at ambient pressure condition (no applied pressure).
Measure and record the output voltage.
Verify that the output is reasonable as per the ratiometric range.
Apply the following pressures one by one:
–1 bar
0 bar
1 bar
2 bar
3 bar
4 bar
5 bar
At each pressure point:
Allow the signal to stabilize.
Measure and record the output voltage on the multimeter.
For a transmitter with range –1 to +5 bar, the output should be approximately:
| Sr. No. | Applied Pressure (bar) | Expected Output |
|---|---|---|
| 1 | –1 | 10% of supply voltage |
| 2 | 0 | 23.33% of supply voltage |
| 3 | 1 | 36.66% of supply voltage |
| 4 | 2 | 50% of supply voltage |
| 5 | 3 | 63.33% of supply voltage |
| 6 | 4 | 76.66% of supply voltage |
| 7 | 5 | 90% of supply voltage |
If supply voltage = 10.00 V
Then:
At –1 bar → Output ≈ 1.00 V
At 2 bar → Output ≈ 5.00 V
At 5 bar → Output ≈ 9.00 V
Output voltage should:
Follow the ratiometric percentage values
Be linear
Be stable at each pressure point
Deviation should be within specified tolerance (as per datasheet).
⚠️ Always measure supply voltage first.
⚠️ Do NOT compare output to fixed voltage values unless supply is perfectly regulated.
⚠️ Output will change if supply voltage changes — this is normal behavior for ratiometric sensors.
If output is shifted → Check supply voltage first.
If output is unstable → Check grounding and shielding.
If output is non-linear → Sensor may be damaged or out of calibration.
This procedure ensures that the pressure transmitter is functioning correctly and producing accurate ratiometric output proportional to applied pressure and supply voltage.
Quasi architecto beatae vitae dicta sunt explicabo. Nemo enim ipsam
voluptatem quia voluptas sit aspernatur aut odit fugit
Can't find the answer you're looking for? Don't worry we're here to solve your softtware problem!
Contact Support