05-December-2025
PL40
55
To define the standard procedure for zero and span calibration of the pressure transducer using a pressure calibrator, ensuring accurate and stable measurement output.
This procedure applies to all pressure transmitters of this model that provide potentiometer-based zero and span adjustment.
Pressure calibrator (hand pump / deadweight / electronic calibrator)
Suitable screwdriver (small, flat)
Spanner or wrench (for loosening plug nut)
Multimeter or loop calibrator (if required for output verification)
✅ The transmitter must be powered ON
✅ The transmitter must be allowed to warm up for at least 30 minutes to reach operating temperature
✅ The transmitter must be connected to a pressure calibrator
❌ Do NOT attempt calibration without a pressure calibrator (this is not a guessing game)
Calibration can only be done using a pressure calibrator.
Zero and Span adjustments are done via internal potentiometers.
The transmitter is sealed using potting compound. The compound reseals itself after adjustment — no drama.
① Plug nut – Must be loosened
② Potting compound access point
③ Zero adjustment potentiometer (±10% approx)
④ Span adjustment potentiometer (±20% approx)
Connect the pressure calibrator to the pressure port of the transmitter.
Power ON the transmitter.
Wait minimum 30 minutes for temperature stabilization.
Loosen and unscrew the plug nut (①).
Insert a screwdriver through the potting compound (②) to reach the potentiometers.
Apply 0% pressure (or LRV) using the pressure calibrator.
Adjust the Zero potentiometer (③) until output matches:
4.00 mA (for 4–20 mA) or
0 V / correct digital value (as applicable)
Zero adjustment range is ±10% approx
Apply 100% pressure (or URV) using the pressure calibrator.
Adjust the Span potentiometer (④) until output matches:
20.00 mA (for 4–20 mA) or
Full-scale output value
Span adjustment range is ±20% approx
Recheck Zero
Recheck Span
Repeat steps 3 and 4 until both zero and span are correct
(Yes, this back-and-forth is normal. Welcome to the real world.)
Remove the screwdriver.
The potting compound will self-close the access hole.
Tighten the plug nut (①) properly.
Check output at:
0%
25%
50%
75%
100%
Ensure output is linear and within tolerance.
Zero and span match specified values
Output is stable and repeatable
No leakage from plug or pressure port
❌ Calibrating without warm-up
❌ Adjusting span before zero
❌ Not looping back to recheck zero after span
❌ Using a screwdriver like you’re tightening a truck wheel
Quasi architecto beatae vitae dicta sunt explicabo. Nemo enim ipsam
voluptatem quia voluptas sit aspernatur aut odit fugit
Can't find the answer you're looking for? Don't worry we're here to solve your softtware problem!
Contact Support